| Recording system |
High resolution themal array printer (8 dots/mm) |
| Recording channel |
1/3-ch (Cardipia200,400) / 6/3-ch(Cardipia 400) |
| Recording paper |
50mm / 63mm x 30mm(CARDIPIA 200)/ 110mm x 30m(CARDIPIA 400) thermal sensitive roll paper |
| Recording ECG leads |
6 standard leads: I, II, III, aVR, aVL, aVF |
| Non-interpretation |
Type 210NV Ype 203NV Type 406NV |
| Parameters |
Type 406PV |
| Sensitivity |
5, 10, 20mm/mV |
| Recording speed |
25,50 mm/sec (CARDIPIA 200) / 5, 10, 12.5, 25, 50 mm/sec (CARDIPIA 400) |
| Recording operation |
Automatic / Manual recording |
| ECG copy |
Reprint for th last patient (CARDIPIA 400) |
| Full disclosure report |
Captured ECG trace for 1min while displaying ECG (CARDIPIA 400) |
| Snap ECG record |
Latest signal save for 8sec while displaying ECG (CARDIPIA 400) |
| ECG acquisition |
Simultaneous 6-lead ECG acquisition |
| Alphanumeric data input |
ID, age, patient name, hospital name and etc with alphanumeric key |
| Frequency response |
Filtered 0.5 ~ 30Hz / Unfiltered 0.05 ~ 150Hz |
| Digital filter |
Drift, muscle, and AC (50/60 Hz) |
| 1 mV calibration |
Internal 1mV signal record for frequency and sensitivity response |
| A/D conversion |
10 bit (CARDIPIA 200) / 12bit (CARDIPIA 400) |
| Sampling rate |
2ms |
| CMPR |
Greater than 100dB |
| Input impedance |
Greater than 50MQ(CARDIPIA 200),Greater than 100MQ(CARDIPIA 400) |
| Performance |
Comply with AAMI requirements |
| Safety |
Class I (Degree of protection against electric shock: Type CF applied parts) |
| Defibrillator protection |
5000 VDC |
| Patient leakage current |
Less than 10 |
| A/C power |
Free voltage AC 100 ~ 240V 10%, 50/60Hz |
| D/C power |
Built-in rechargeable battery (+24V, 600mA, NiCd) |
| Digital Interface(Option) |
RS232 to connect external apparatus(CARDIPIA 400) |
| Heart rate(Option) |
30 ~ 200bpm |
| PC network(Option) |
ECG date transmission to PC, ECG save on PC, ECG record to PC printer |
| Real time clock(Option) |
Date, time (CARDIPIA 400) |
| Arrhythmia detection(Option) |
Automatic record mode(CARDIPIA 400) |
| Built-in Memory(Option) |
Up to 70 patients (CARDIPIA 400) |
| CARDIPIA 200 LCD(Standard) |
Character LCD (16 X 2 Lines) viewing area: 65 X 15mm |
| CARDIPIA 400 LCD |
Standard : Graphic LCD(64 x 128dots, 2.5"LCD,1-ch display), Viewing area: 60 x 32.5mm |
| Option : Graphic LCD(240 x 320dots, 4"LCD,6-ch dispaly), Viewing area: 57 x 75mm |
| Auto Power off |
After 5 minutes of complete inactivity |
| Dimension |
313 X 252 X 85mm |
| Weight |
2.5kg including rechargeable battery |
| Temperature |
Operating : 10 ~ 40 , Storage : -40 ~ -70 |
| Relative humidity |
Operating : 30 ~ 70%, Storage : 10 ~ 80% |
| Atmosphere |
700 ~ 1060 hPa |